DSHD-3536D Fully Automatic Cleveland Open Cup Flash Point Tester

Automatic temperature control, flash point capture, result print.
Could control by PC.
Test fire point as well.

  • Range: 80-400℃
  • Standards: ASTM D92, ISO 2592, GB/T 3536, T0611
Quantity:

Product Description

Summary
This instrument is designed and made as per GB/T3536-2008 Petroleum Products - Determination of Flash and Fire Points - Cleveland open cup method and ASTM D 92 Standard Test Method for Flash and Fire Points by Cleveland Open Cup Tester. It is applicable to all petroleum products with flash points above 79℃ and below 400℃ except fuel oils.
 

Technical Data Sheet

I。 Main technical features
1. This instrument is a kind of fully-automatic instrument wich can do determination for both flash point and fire point.
2. This instrument adopts technology of single chip microcomputer and LCD screen.Full English man-machine dialog interface. It can preset the parameters of expected flash point,sample mark number, atmospheric pressure, test date,etc.It has menu to prompt and input function of guide type.
3. This instrument adopts desktop structure and advanced design.The man-machine interface make it easy to use.And the test result is accurate.It can be widely used in railway,aviation,electric power, petroleum industry and colleges,scientific research institutes,measurement departments to do determination of flash point and fire point for petroleum products.
 

 

II. Main technical specifications
SN。  Item  Parameter
1  Power supply  AC 220V(-10%~+5%), 50 Hz
2  Temperature measurement  Temperature measurement
2.1  Full scale  Room temperature~400℃
2.2  Repeatability  ≤4℃
2.3  Reproducibility  ≤8℃
2.4   Resolution  0.1 ℃
2.5  Accuracy  0.5%
3  Temperature rising speed  Conform to GB/T3536
4  Flame application  Electronic ignition ; gas flame is about 3。2~4。8 mm in length
5  Ambient temperature  10℃~40℃
6  Relative humidity  ≤80%;
7  Total power consumption  Not more than 500 W
8  Dimension
8。1  When test arm doesn’t raise up  520mm×360mm×310mm

 When test arm raised up  520mm×360mm×420mm